Table 2 Recommended reconstruction parameters IQ-SPECT

From: EANM procedural guidelines for myocardial perfusion scintigraphy using cardiac-centered gamma cameras

99mTc data

201Tl data

Gated reconstruction parameters:

Gated reconstruction parameters:

 12 iterations

 12 iterations

 1 subset

 1 subset

 10-mm Gaussian smooth (adjust as needed)

 7–10-mm Gaussian smooth (adjust as needed). No scatter correction.

Approximate reconstruction times:

Approximate reconstruction times:

 8 time bins → 1.5 min

 8 time bins → 1.5 min

 16 time bins → 3.0 min

 16 time bins → 3.0 min

Non-gated AC and No AC reconstruction parameters:

Non-gated AC and No AC reconstruction parameters:

 10 iterations

 10 iterations

 3 subsets

 3 subsets

 10-mm Gaussian smooth (adjust as needed)

 7–10-mm Gaussian smooth (adjust as needed). No scatter correction

 Approximate reconstruction time:

 Approximate reconstruction time:

 Less than 1 min for single non-gated data set, AC and NC

 Less than 1 min for single non-gated data set, AC and NC