Skip to main content

Table 1 Quantitative measures in a liver sphere with a diameter of 3 cm. Data represents mean value and range (* indicates significantly different value, p<0.01, for DDG compared with either static or external device-gated acquisitions)

From: Quantitative comparison of data-driven gating and external hardware gating for 18F-FDG PET-MRI in patients with esophageal tumors

 

Static acquisition

Gated acquisition

Measure

First 0–3 min

Last 4–6 min

External device

DDG

SUVmean

2.37 (1.59–2.73)

2.31 (1.62–2.67)

2.27 (1.54–2.65)

2.27 (1.56–2.63)

SUVSD

0.53 (0.37–0.83)

0.52 (0.37–0.74)

0.53 (0.35–0.76)

0.37 (0.26–0.61)*

Noise level

0.225 (0.15–0.33)

0.223 (0.15–0.29)

0.235 (0.15–0.32)

0.165 (0.12–0.21)*

SNR

4.61 (3.07–6.51)

4.73 (3.05–6.45)

4.46 (2.96–6.46)

6.32 (4.67–8.50)*